The DMA 35N portable density meter measures
the density of liquids in g/cm3 or kg/cm3 according to the U-tube principle.
The sample is inserted into the measuring cell using the built-in pump or a syringe. A temperature sensor measures the temperature of the sample in the measuring cell. The temperature is displayed on the read-out panel and at the same time used in the automatic calculation of density measurement as a function of temperature. Once the calculator is activated, the measured density value is immediately converted into density at the reference temperature (15°C). The DMA density meter has a storage capacity of 1024 measurement results,
including temperature values.
|
![]() |
SPECIFICATIONS :
|
The TP7-C employs the proven RTD design that has been used in the TP 7 and TP 8 for many years. A sealed industrial quality overlay provides a user interface that is easy to use with gloves. A polypropylene blend enclosure protects the circuit board and large LCD from penetration by impact, water and reactive liquids.
The new TP 7-C circuit board is an evolutionary step up from our highly accurate, reliable and successful TL1 laboratory thermometer. The Power Button’s primary function powers the instrument for intervals of about 20 minutes since the last button was accessed. The Power Button can also be used to conserve power and clear the memory, or to make adjustment while in calibration mode. A simple menu operation is displayed by holding the Function Button, and alternately functions to allow adjustments in the calibration mode. Arrows on the left side of the display show the direction of the temperature reading and whether stability has been reached. At the user’s discretion stabilized temperatures can be logged at numerous liquid levels for a running average and later displayed for the user’s documenting purposes. But this feature never interferes with simply getting an accurate temperature reading.
To endure the environment and be intrinsically safe, the TP 7-C is manufactured of materials, which are both immune to petrochemicals and are non-sparking. The enclosure is made of a polypropylene blend. The probe assembly is constructed using non-stick cable and stainless steel sensor components..
Operational Attributes:
Easily replaceable AA Batteries, provides an estimated 200 hours operation. Circuit
logic automatically indicates low battery condition, automatically shuts off
after twenty minutes, shows temperature trend and stabilization, displays error
codes for failure determination.
The low power backlight for night operation
is photo sensor controlled for convenience and battery conservation.
In nighttime
conditions the backlight illuminates the display.
Celsius or Fahrenheit units
with C/F indication can be easily selected from the Function Button.
User
Manual explains intuitive calibration procedure that can be done through the
external faceplate buttons.
The TP-7C is
an Intrinsically Safe Thermometer :
Class I, Div I, Groups ABCD Hazardous
locations
ATEX II 2G Ex ia IIB T4
SPECIFICATIONS :
|
|
|||
ATEX II 2G Ex ia II B T4 EN 50014 . EN 50020 |
This portable probe performs 3 tasks in a single operation:
|
![]() |
Measurement precision: | + 2 mm | Temperature resolution: | + 0.1° C |
Dip tape length: | 15 or 30 m | Time constant: | approx. 10 seconds |
Level measurement precision: | +1.5 mm/30m |
Measurement precision: | + 0.1° C |
Dip tape coating: | TEFZEL® | Battery power level indicator: | accoustic + display |
Probe diameter: | 23 mm | Temperature display: | in °C or °F |
Probe material: | Inox 316, Viton® | ||
Detectable minimum: | 4 mm | Battery type: | 9V - 6F22 |
Ambient operating temperature: | -20° to +50° C | Intrinsic safety: | Kema/ATEX |
Measurable temperature: | -40° to + 90° C | Weight: | 4.5 KG |
LABORATORY
THERMOMETER
THERMOPROBE TL1A
The TL1-A intrinsically safe thermometer is intended to be used anywhere a precision glass stem thermometer or other type thermometer may be used.The TL1-A is an improved version of the original ThermoProbe model TL1. The TL1-A incorporates glass thermometer simplicity and reliability with the ease of a digital display. Like precision mercury thermometers, the TL1-A has accuracy, repeatability and long-term stability. But unlike those thermometers there will never be problems with mercury separation or hazardous material cleanup. Different from most electronic thermometers, the "Cordless" TL1-A has no wires to get in the way or break. Battery replacement is easily done through a sealed cover. Encoded calibration is unaffected by temperature, vibration, or battery removal. Low battery display, and failure mode indications protect the user from false data. |
|
4 Point NIST Traceable Calibration and a one year warranty included.
KEY FEATURES
Auto-off after 20 minutes to save battery life
Minimum reading
Maximum reading
Averaging
Easy °F to °C switching
Display arrows indicate increasing or decreasing temperature activity
Only 3 onces, 93 grams
Display |
LCD with 0.25 inch characters Trend Indicating Arrows Low Battery & Fault indication |
Construction |
Stainless Steel temperature sensor; Standard Sensor length 8" (20cm) Standard diameter 1/4" (6.35mm) with reduced diameter tip of 3/16" (4.76mm) Sensor mount acetyl plastic Electronics and display in 3.5" x 1.2" anodized Aluminum enclosure Molded Silicone seal. Highly durable display window, graphics and membrane switches. |
Temperature |
Resolution 0.01 Degrees Calibrated Accuracy: ±0.1°F from 14 to 320°F ±0.06°C from -10 to 160°C NIST Traceable Report of Test |
Electronics |
Microcontroller and precision 20 bit A/D Converter FLASH memory Recommended Temperature Range For best Accuracy 32 to 112°F, 0 to 44°C 100 Ohm, Class A Platinum RTD Immersion depth 4", 11cm |
Sensor |
100 Ohm, Class A 1/3 DIN RTD Immersion depth 4", 11cm |
Battery |
20mm Coin Size 3V Lithium, Duracell DL2032(IEC CR2032) Over 100 hours Service Life at 70°F, 21°C Battery Operating Range -4 to 130°F, -20 to 54°C Note: Battery may not provide adequate power if ambient temperature is below -4°F, -20°C or above 130°F, 54°C |
User Fonctions |
Auto Off after 20 minutes or continuous operation Lowest Reading, Highest Reading, Average Reading Switchable in Units of °F or °C Display tenths or hundredths resolution |
|